Misuratore di rigidità AC/DC fino a 200 VA, di isolamento e di continuità.
AC WITHSTANDING
Output voltage range: from 0.100 to 5.000 kV AC
Output voltage resolution: 2 V/step
Output voltage accuracy: ±1% of setting +5 V (no load)
Maximum rated load: 200 VA (5 kV / 40 mA)
Maximum rated current:
- 40 mA (0.5 kV < V ? 5 kV)
- 10 mA (0.1 kV ? V ? 0.5 kV)
DC WITHSTANDING
Output voltage range: from 0.100 kV to 6.000 kV DC
Output voltage resolution: 2 V/step
Output voltage accuracy: ±(1% of setting +5 V (no load)
Maximum rated load: 50 W (5 kV / 10 mA)
Maximum rated current:
- 10 mA (0.5 kV < V ? 6 kV)
- 2 mA (0.1 kV ? V ? 0.5 kV)
INSULATION RESISITANCE
Output voltage range: from 50 to 1000 V DC
Output voltage resolution: 50 V/step
Output voltage accuracy: ±1% of setting +5 V (no load)
Resistance measurement range: from 1 to 9500 MΩ
GROUND BOND
Output current range: from 3.00 to 30.00 A AC
Output current resolution: 0.01 A
Output current accuracy:
- ±1% of setting +0.2 A (3 A ? I ? 8 A)
- ±1% of setting +0.05 A (8 A < I ? 30 A)
240 x 64 Ice Blue Dot matrix LCD
Manual/Auto mode, function key for quick selecting, high intensity flash for caution & status indication
Safety INTERLOCK function, zero crossing turn-on operation, controllable ramp-up time
True RMS current measurement, high resolution (1 µA for measuring current, 2 V for setting voltage)
PWM amplifier to enhance the power efficiency and reliable testing
Max. 100 memory block for test condition (step) setting; each step can be named individually
Remote terminal on the front panel for “start” and “stop” control
Interface : RS-232C, USB device, Signal I/O and GPIB (optional)